Your device contains calibration D/A converters (calDACs) that are used for fine-tuning the 
analog circuitry. The calDACs must be programmed (loaded) with certain numbers called 
calibration constants. These constants are stored in nonvolatile memory (EEPROM) on your 
device. To achieve specification accuracy, you should perform an internal calibration of your 
device just before a measurement session but after your computer and the device have been 
powered on and allowed to warm up for at least 15 minutes. Frequent calibration produces the 
most stable and repeatable measurement performance.
Before the device is shipped from the factory, an external calibration is performed and the 
EEPROM contains calibration constants that NI-DAQ automatically loads into the calDACs 
as needed. The value of the onboard reference voltage is also stored in the EEPROM, and this 
value is used when you subsequently perform a self-calibration. The calibration constants are 
recalculated and stored in the EEPROM when a self-calibration is performed. When you 
perform an external calibration, NI-DAQ recalculates the value of the onboard reference 
voltage and then performs a self-calibration. This new onboard reference value is used for all 
subsequent self-calibration operations. If a mistake is made when performing an external 
calibration, you can restore the board’s factory calibration so that the board is not unusable.
Performing Self-Calibration of the Board
Set operation to ND_SELF_CALIBRATE to perform self-calibration of your device.
Example:
You want to perform self-calibration of your device and you want to store the new set of 
calibration constants in the EEPROM. You should make the following call:
Calibrate_DSA (deviceNumber, ND_SELF_CALIBRATE, 0.0)
Performing External Calibration of the Board
Set operation to ND_EXTERNAL_CALIBRATE to externally calibrate your device. The value 
of the internal reference voltage will be recalculated and the board will be self-calibrated 
using the new reference value.
Before calling the 
Calibrate_DSA function, connect the output of your reference voltage to 
analog input channel 0.
Example:
You want to externally calibrate your device using an external reference voltage source with 
a precise 7.0500 V reference, and you want to store the new set of calibration constants in the 
EEPROM. You should make the following call:
Calibrate_DSA (deviceNumber, ND_EXTERNAL_CALIBRATE, 7.0500)