User's Manual
Chapter 2 Function Reference — SCXI_Cal_Constants
NI-DAQ FRM for PC Compatibles 2-326
©
 National Instruments Corporation
milliamperes. If opCode = 3, you should pass your second calibration constant in calConst2 
for NI-DAQ to store in calibrationArea.
Note C Programmers—calConst1 and calConst2 are pass-by-reference parameters.
Using This Function
Analog Input Calibration
When you call SCXI_Scale to scale binary analog input data, NI-DAQ uses the binary offset 
and gain adjust calibration constants loaded for the given module, channel, and gain setting 
to scale the data to voltage or frequency. Refer to the 
SCXI_Scale function description for 
the equations used.
By default, NI-DAQ loads calibration constants for the SCXI-1122, SCXI-1126, and 
SCXI-1141 from the module EEPROM (see the EEPROM Organization section later in 
this function for more information). The SCXI-1141 has only gain adjust constants in the 
EEPROM and does not have binary zero offset in the EEPROM. All other analog input 
modules have no calibration constants by default; NI-DAQ assumes no binary offset and ideal 
gain settings for those modules unless you use the following procedure to store calibration 
constants for your module.
You can determine calibration constants based specifically on your application setup, which 
includes your type of DAQ device, your DAQ device settings, and your cable assembly, all 
combined with your SCXI module and its configuration settings.
Note NI-DAQ stores constants in a table for each SCXI module gain setting. If your 
module has independent gains on each channel, NI-DAQ stores constants for each 
channel at each gain setting. When you use the following procedure, you are also 
calibrating for your DAQ device settings, so you must use the same DAQ device 
settings whenever you use the new calibration constants. The SCXI-1122, 
SCXI-1126, and SCXI-1141 factory-set EEPROM constants apply only to the 
SCXI-1122, SCXI-1126, and SCXI-1141 amplifiers, respectively, so you can use 
those with any DAQ device setup.
To perform a two-point analog input calibration, perform the following steps:
1. If you are using an AT-MIO-16F-5, AT-MIO-64F-5, or AT-MIO-16X device, you should 
calibrate your ADC first using the 
MIO_Calibrate function.
2. Make sure the SCXI gain is set to the gain you will be using in your application. If 
you are using an SCXI-1100, SCXI-1122, SCXI-1126, or SCXI-1141, you can use the 
SCXI_Set_Gain function, because those modules have software-programmable gain. 
For other analog input modules, you need to set gain jumpers or DIP switches 
appropriately.
3. Use 
SCXI_Single_Chan_Setup to program the module for a single-channel operation 
(as opposed to a channel scanning operation).