Universal Analog Input Module User Guide and Specifications

© National Instruments Corporation 11 NI USB-9219 User Guide and Specifications
Connecting Wires to the NI USB-9219 Connectors
Use a flathead screwdriver with a blade smaller than 2.3 × 1.0 mm (0.09 ×
0.04 in.) to connect wires to the detachable spring-terminal connectors.
Insert the screwdriver into a spring clamp activation slot and press a wire
into the corresponding connector terminal, then remove the screwdriver to
clamp the wire into the terminal. Refer to the Specifications section for
more information about spring-terminal wiring. Refer to Figure 7 for an
illustration of connecting wires to the NI USB-9219.
Figure 7. Connecting Wires to the NI USB-9219 Connectors
Table 4. Terminal Assignments
Mode
Terminal
1 2 3 4 5 6
Vo l t a g e T+ T HI LO
Current T+ T HI LO
4-Wire Resistance T+ T EX+ HI EX LO
2-Wire Resistance T+ T HI LO
Thermocouple T+ T HI LO
4-Wire RTD T+ T EX+ HI EX LO
3-Wire RTD T+ T EX+ EX LO
Quarter-Bridge T+ T HI LO
Half-Bridge T+ T EX+ HI EX
Full-Bridge T+ T EX+ HI EX LO
Digital In T+ T HI LO
Open Contact T+ T HI LO