Device User Guide and Specifications
NI USB-6008/6009 User Guide and Specifications 26 ni.com
Timing accuracy .....................................100 ppm of actual sample rate
Input range
Single-ended....................................±10 V
Differential ......................................±20 V
1
, ±10 V, ±5 V, ±4 V,
±2.5 V, ±2 V, ±1.25 V, ±1 V
Working voltage .....................................±10 V
Input impedance .....................................144 kΩ
Overvoltage protection ...........................±35
Trigger source.........................................Software or external digital
trigger
System noise
2
Single-ended
±10 V range..............................5 mVrms
Differential
±20 V range .............................5 mVrms
±1 V range................................0.5 mVrms
Absolute accuracy at full scale, single ended
Absolute accuracy at full scale, differential
3
1
±20 V means that |AI+ – (AI–)| <= 20 V. However, AI+ and AI– must both be within ±10 V of GND. Refer to the Connecting
Differential Voltage Signals for more information.
2
System noise measured at maximum sample rate.
Range
Typical at 25 °C
(mV)
Maximum over Temperature
(mV)
±10 14.7 138
3
Input voltages may not exceed the working voltage range.
Range
Typical at 25 °C
(mV)
Maximum over Temperature
(mV)
±20 14.7 138
±10 7.73 84.8
±5 4.28 58.4
±4 3.59 53.1
±2.5 2.56 45.1