- NI 6013/6014 User Manual

Chapter 4 Connecting Signals
NI 6013/6014 User Manual 4-6 ni.com
Analog Input Signal Overview
The AI signals for the NI 6013/6014 are ACH<0..15>, AISENSE, and
AIGND. Connection of these AI signals to the device depends on the type
of input signal source and the configuration of the AI channels you are
using. This section provides an overview of the different types of signal
sources and AI configuration modes. More specific signal connection
informationisprovidedintheConnecting Analog Input Signals section.
DIO<0..7> DIO V
CC
+0.5 10 at (V
CC
0.4) 24 at
0.4
1.1 1.5 kpd
SCANCLK DO 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
EXTSTROBE* DO 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI0/TRIG1 DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI1/TRIG2 DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI2/CONVERT* DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI3/GPCTR1_SOURCE DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI4/GPCTR1_GATE DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
GPCTR1_OUT DO 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI5/UPDATE* DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI6/WFTRIG DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI7/STARTSCAN DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI8/GPCTR0_SOURCE DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
PFI9/GPCTR0_GATE DIO V
CC
+0.5 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
GPCTR0_OUT DO 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
FREQ_OUT DO 3.5 at (V
CC
0.4) 5at0.4 1.5 50 kpu
pd = pull down
pu = pull up
DO = Digital Output
The tolerance on the 50 kpull-up resistors is very large. Actual value may range between 17 and 100 k.
Table 4-3. I/O Signal Summary for the NI 6013/6014 (Continued)
Signal Name
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA
at V)
Rise
Time
(ns)
Bias