- NI 6013/6014 User Manual
Chapter 4 Connecting Signals
© National Instruments Corporation 4-5 NI 6013/6014 User Manual
PFI8/GPCTR0_SOURCE DGND Input
Output
PFI8/Counter 0 Source—As an input, this signal is a PFI.
As an output, this signal is the GPCTR0_SOURCE signal.
This signal reflects the actual source connected to the
general-purpose counter 0.
PFI9/GPCTR0_GATE DGND Input
Output
PFI9/Counter 0 Gate—As an input, this signal is a PFI.
As an output, this signal is the GPCTR0_GATE signal.
This signal reflects the actual gate signal connected to the
general-purpose counter 0.
GPCTR0_OUT DGND Output Counter 0 Output—This output is from the general-purpose
counter 0 output.
FREQ_OUT DGND Output Frequency Output—This output is from the frequency
generator output.
*
Indicates that the signal is active low.
1
Not available on the NI 6013.
Table 4-3. I/O Signal Summary for the NI 6013/6014
Signal Name
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA
at V)
Rise
Time
(ns)
Bias
ACH<0..15> AI 100 GΩ
in
parallel
with
100 pF
25/15 — — — ±200 pA
AISENSE AI 100 GΩ
in
parallel
with
100 pF
25/15 — — — ±200 pA
AIGND AO — — — — — —
DAC0OUT
(NI 6014only)
AO 0.1 Ω Short-circuit
to ground
5at10 5at
–10
4
V/µs
—
DAC1OUT
(NI 6014 only)
AO 0.1 Ω Short-circuit
to ground
5at10 5at
–10
4
V/µs
—
AOGND AO — — — — — —
DGND DO — — — — — —
V
CC
DO 0.1 Ω Short-circuit
to ground
1A fused — — —
Table 4-2. Signal Descriptions for I/O Connector Pins (Continued)
Signal Name Reference Direction Description