user manual
USB-9201/9221 User Guide and Specifications 6 ni.com
Mounting the USB-9201/9221 to a Panel
Threaded inserts are located in the USB-9201/9221 for mounting it to a
panel. Refer to Figure 6 for dimensions.
Figure 6. Module Dimensions
Connecting the USB-9201/9221 to a Computer
Plug one end of the USB cable into the USB-9201/9221 and the other end
into an available USB port on the computer. Refer to the NI-DAQmx for
USB Devices Getting Started Guide that shipped with your device, and is
also accessible from Start»All Programs»National Instruments»
NI-DAQ, for more information.
LED Indicator
The LED indicator indicates device status.
Table 1. LED State/Device Status
LED State Device Status
Not lit Device not connected or in suspend.
On, not blinking Device connected, but no module installed.
Single-blink Operating normally.
Threaded Insert
M3 x 0.5
8.5 mm (0.34 in.) Max Depth
72.2 mm
(2.84 in.)
85.7 mm
(3.37 in.)
76.1 mm
(3.00 in.)