user manual
Table Of Contents
- NI PXI/PCI-5124 Specifications
- Contents
- Vertical
- Analog Input (Channel 0 and Channel 1)
- Figure 1. NI 5124 Frequency Response (Typical)
- Figure 2. NI 5124 Dynamic Performance, 50 W , 1 Vpk-pk Range, 262,144 Point FFT (Typical)
- Figure 3. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Noise Filter Enabled, 1 MW Input Impedance
- Figure 4. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Full Bandwidth, 50 W Input Impedance
- Analog Input (Channel 0 and Channel 1)
- Horizontal
- Trigger
- PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connector)
- TClk Specifications
- Waveform Specifications
- Calibration
- Power
- Software
- Environment
- Safety, Electromagnetic Compatibility, and CE Compliance
- Physical
- Where to Go for Support

© National Instruments Corporation 9 NI PXI/PCI-5124 Specifications
Specification Val ue Comments
RMS Noise Range (V
pk-pk
) 50 Ω 1MΩ Noise filter on
50 Ω
terminator
connected
to input
0.2 NI PXI-5124:
94 μV
rms
(0.047% FS)
NI PCI-5124:
106 μV
rms
(0.053% FS)
NI PXI-5124:
104 μV
rms
(0.052% FS)
NI PCI-5124:
116 μV
rms
(0.058% FS)
0.4 188 μV
rms
(0.047% FS)
192 μV
rms
(0.048% FS)
1 470 μV
rms
(0.047% FS)
480 μV
rms
(0.048% FS)
2 940 μV
rms
(0.047% FS)
960 μV
rms
(0.048% FS)
4 1.88 mV
rms
(0.047% FS)
1.92 mV
rms
(0.048% FS)
10 4.7 mV
rms
(0.047% FS)
4.8 mV
rms
(0.048% FS)
20
(1 MΩ only)
— 9.4 mV
rms
(0.047% FS)