user manual
Table Of Contents
- NI PXI/PCI-5124 Specifications
- Contents
- Vertical
- Analog Input (Channel 0 and Channel 1)
- Figure 1. NI 5124 Frequency Response (Typical)
- Figure 2. NI 5124 Dynamic Performance, 50 W , 1 Vpk-pk Range, 262,144 Point FFT (Typical)
- Figure 3. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Noise Filter Enabled, 1 MW Input Impedance
- Figure 4. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Full Bandwidth, 50 W Input Impedance
- Analog Input (Channel 0 and Channel 1)
- Horizontal
- Trigger
- PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connector)
- TClk Specifications
- Waveform Specifications
- Calibration
- Power
- Software
- Environment
- Safety, Electromagnetic Compatibility, and CE Compliance
- Physical
- Where to Go for Support

© National Instruments Corporation 11 NI PXI/PCI-5124 Specifications
RMS Noise Range (V
pk-pk
) 50 Ω 1MΩ Filters off
50 Ω
terminator
connected
to input
0.2 NI PXI-5124:
114 μV
rms
(0.057% FS)
NI PCI-5124:
128 μV
rms
(0.064% FS)
NI PXI-5124:
164 μV
rms
(0.082% FS)
NI PCI-5124:
184 μV
rms
(0.092% FS)
0.4 204 μV
rms
(0.051% FS)
264 μV
rms
(0.066% FS)
1 510 μV
rms
(0.051% FS)
550 μV
rms
(0.055% FS)
2 1.02 mV
rms
(0.051% FS)
1.08 mV
rms
(0.054% FS)
4 2.08 mV
rms
(0.052% FS)
2.6 mV
rms
(0.065% FS)
10 5.2 mV
rms
(0.052% FS)
5.5 mV
rms
(0.055% FS)
20 — 10.6 mV
rms
(0.053% FS)
Specification Val ue Comments