Deterministic Ethernet Expansion Chassis User's Guide
NI 9144 User Guide and Specifications 38 ni.com
Note LSB weight is referred to as Gain in the object dictionary.
The calibration data is stored in a U32 array, though each Offset field
(subindex 1, 3, 5, and so on) should be interpreted as a signed value.
Use the calibration coefficients with the following equation to generate
correct data:
R
raw
(bits) = data returned by the NI 9217 in bits
R
corrected
= calibrated resistance reading
NI 9219
Table 31. NI 9217 Calibration Coefficients
Coefficient Representation Units
LSB Weight Unsigned pΩ/LSB
Offset Signed μΩ
Table 32. NI 9219 Vendor Configuration Extensions
Index Sub Type R/W Description
0x2001 0 ARR:U32 — Command List = 33
1 R Command Count = 1..32, default = 32
2..33 R/W Configuration Command
0x2002 1 ARR:U32 R Error Status
0x2005 0 U32 R/W ADC Format
R
corrected
R
raw
()R
raw
bits()
∗
LSB
weight
pΩ
bits
---------
⎝⎠
⎛⎞
∗
10
12–
Ω
pΩ
-------
⎝⎠
⎛⎞
⎩⎭
⎨⎬
⎧⎫
Offset μΩ()
∗
10
6–
Ω
μΩ
--------
⎝⎠
⎛⎞
–=