User manual

30
2. Launch the application.
3. Enable channels 1 and 2 of the MSO by clicking the square icons next to the channel names. Set the vertical
settings for each channel to 500 mV/div.
Figure 3.8. Disabled MSO Inputs (left) and Enabled Inputs (right)
4. Configure the MSO to have Normal record acquisition and set the horizontal timing to 10 ms/div (both shown in
Figure 3.9). Configure the Trigger Type to Edge, the Channel Source to 1, and the Edge Detection to Rising (as
shown in Figure 3.10). Finally, set the Trigger Level to 0 V (as shown in Figure 3.11).
Figure 3.9. Configuring MSO Record Acquisition Mode and Horizontal Timing Settings
Figure 3.10. Configuring MSO Trigger Type, Source, and Detection Edge