User manual

21
Figure 2.17. AC Voltage Measurements of the Corrected Circuit Model
Table 2.4 shows the simulated measurements of the corrected circuit and your actual measurements side by side. The
measurements are now closer to what you expected to see. If you would like to verify these results for yourself, please
use the VirtualBench Section 2_DMM and FGEN Resistor Network Corrected.ms13 file.
Simulated Voltage
(Corrected)
Measured Voltage
Voltage Source
3.898 V
3.909 V
DC signal source
1.378 V
rms
1.381 V
rms
AC signal source
Table 2.4. Corrected Circuit Voltage Measurement Comparisons
To conclude, when using any signal source, always check for any output impedance when modeling the source.