Datasheet

Power input connector IEC C13 power connector
Power disconnect The AC power cable provides main power
disconnect. Do not position the equipment so
that it is difficult to disconnect the power
cable. Depressing the front panel power button
does not inhibit the internal power supply.
Calibration
Calibration cycle (digital multimeter,
mixed signal oscilloscope, function generator,
DC power supply)
1 year
Specified temperature T
cal
± 5 °C
Warmup time 30 minutes
Physical Characteristics
Dimensions
Enclosure 25.40 cm × 19.05 cm × 7.77 cm
(10.00 in. × 7.50 in. × 3.06 in.)
Enclosure with connectors and antenna 25.40 cm × 23.37 cm × 14.40 cm
(10.00 in. × 9.20 in. × 5.67 in.)
Note Use the VirtualBench instrument in a horizontal orientation. Allow at least
10.16 cm (4.0 in.) of clearance in front and behind the VirtualBench instrument for
USB, power, and common connector cabling.
Weight 2.05 kg (4 lb 8.3 oz)
Connectivity
Mixed signal oscilloscope BNC
Logic analyzer 2x20 shrouded IDC header
External trigger BNC
Function generator BNC
NI VirtualBench VB-8012 Specifications | © National Instruments | 13