Chapter 2 Function Reference — AO_Calibrate
NI-DAQ FRM for PC Compatibles 2-26
©
 National Instruments Corporation
Using This Function
When NI-DAQ initializes the AT-AO-6/10, the DAC calibration constants stored in 
EEPROMloc 1 (user calibration area 1) provide the gain and offset values used to ensure 
proper device operation. In other words, 
Init_DA_Brds performs the equivalent of calling 
AO_Calibrate with operation set to 1 and EEPROMloc set to 1. When the AT-AO-6/10 
leaves the factory, EEPROMloc 1 contains a copy of the calibration constants stored in 
EEPROMloc 5, the factory area.
A calibration procedure performed in bipolar mode is not valid for unipolar and vice versa. 
See the calibration chapter of the AT-AO-6/10 User Manual for more information regarding 
calibrating the device.