Welding System User Manual
Chapter 2 Function Reference — SCAN_Start
NI-DAQ FRM for PC Compatibles 2-310
©
 National Instruments Corporation
2 through 2
24
*
 (total number of channels scanned) or 2
32
 – 1, whichever is less 
for E Series and DSA devices. For PCI-611X devices, count must be EVEN.
count must be an integer multiple of the total number of channels scanned. count refers to 
the total number of A/D conversions to be performed; therefore, the number of samples 
acquired from each channel is equal to count divided by the total number of channels 
scanned. This number is also the total number of scans. For the E Series devices, the total 
number of scans must be at least 2. If you do not use external multiplexer (AMUX-64T) 
devices, the total number of channels scanned is equal to the value of numChans (see 
Scan_Setup). 
If you use one or more external multiplexer devices with any MIO or AI device except the 
MIO-64, the total number of channels scanned equals (four-to-one multiplexer) 
*
 (number of 
onboard channels scanned) 
*
 (scanned) 
*
 (number of external multiplexer devices), or the 
total number of channels scanned equals (4) 
*
 (numChans) 
*
 (num_mux_brds). For example, 
if you use one AMUX-64T and scan eight onboard channels, the total number of channels 
scanned equals (4) 
*
 (8) 
*
 (1) = 32.
If you use one or more external multiplexer devices (AMUX-64Ts) with the MIO-64, the total 
number of channels scanned equals (4) 
*
 (numChans1) 
*
 (num_mux_brds) + numChans2, 
where:
• 4 represents a four-to-one multiplexer.
• numChans1 is the number of onboard channels (of an MIO or AI connector, the first 
connector) scanned. 
Range: 0 through 7 differential, 0 through 15 single-ended.
• num_mux_brds is the number of external multiplexer devices.
• numChans2 is the number of onboard channels (of an analog connector, the second 
connector) scanned.
Range: 0 through 23 differential, 0 through 48 single-ended.
If you use SCXI, the total number of channels scanned is the total number of channels 
specified in the 
SCXI_SCAN_Setup call.
sampTimebase selects the clock frequency that indicates the timebase, or resolution, to be 
used for the sample-interval counter. The sample-interval counter controls the time that 
elapses between acquisition of samples within a scan sequence.
sampTimebase has the following possible values:
–3: 20 MHz clock used as a timebase (50 ns resolution) (E Series only).
–1: 5 MHz clock used as timebase (200 ns resolution) (AT-MIO-16F-5, 
AT-MIO-64F-5, and AT-MIO-16X only).










