Certification

Cascade Technical Sciences, Inc. Certification 9730 Data
The objective of this test program was to subject customer provided test hardware to environmental simulation in compliance with c ustomer stated specification,
including any authorized modification, deviations or concessions to the original requirements. The hardware consisted of items identified in the appropriate sections of
this report. In addition to test hardware identification, each section contains information that describes the associated test setup and performance and the resulting data.
CascadeTEK, Inc measuring instruments used in testing were calibrated according to the requirements of ANSI/NCSL Z540-1-1944, ISO/IEC 17025-2005 and are
NIST traceable. Calibration records are on file and available for inspection by request. Because the test methods are well established and are qualitative or semi-
quantitative in nature, CascadeTEK, Inc does not apply measurement uncertainty unless obligated by contract. Measured value related to the corresponding tolerance
requirement is use d to decide whether a test meets the requirements of the specification. Any test hardware operational setups and resulting evaluations or inspections
performed by the customer are not included in this report, unless they were explicitly requested. While observations and/or specification compliance statements may
be reported, no interpretations or opinions regarding customer product performance are intended. Unless otherwise indicated in the a ppropriate report section, all
contract obligations were met and the test objective achieved.
2
6. Dust Test per Reference (d) Paragraph 13.4, (IP-6X)m Category 2 and (f) Item 6,
samples were exposed to 8 hours of dust exposure with a concentration of 2 kg/m³
using talcum powder as dust. (Foam)
7. Temporary Immersion Test per Reference (d) Paragraph 14.2.7, (IP-X7) and (f) Item
7, samples were immersed in 1 meter of water for a period of 30 minutes. (Foam)
Testing was done in accordance with the above references as evidenced and reported in the
accompanying data. The test samples were returned to the customer’s facility for
evaluation.
The original of this report is on file at Cascade Technical Sciences, Inc. under the above
referenced certification number for review by authorized personnel. The results of the testing
reported herein relate only to the actual item tested.
Respectfully submitted,
David Bowles
Quality Manager
Cascade Technical Sciences, Inc.
This test certification shall not be reproduced, except in full, without written authorization from Cascade Technical Sciences.
Total number of pages in this document is 124.