The mScan™ device comply with IEC 60601-1 and IEC 60601-1-2. The mScan™ device is tested
to meet requirements of IEC 60601-1 and IEC 60601-1-2.
The recommended wireless security measures include AES encryption.
mScan™ device FCC ID is 2AWC6-1301.
Calibration Requirements
No calibration is required for the mScan™ .
Cleaning
Caution: To avoid transmission of infection, wipe down the mScan™ between each
patient use. The mScan™ may be wiped down with alcohol wipes. Under no
conditions should the mScan™ be submerged in water or any other fluid. Care should
be taken to prevent fluids from coming into contact with any part of the mScan™ .
Storage, Maintenance and Repairs
The mScan™ should be stored in a dry, clean, dust-free environment away from
harmful chemicals, where the temperature never exceeds 95 degrees Fahrenheit.
If the mScan™ appears to be malfunctioning, contact Myolex technical support for assistance
by calling 888-382-8824 x10 or emailing support@myolex.com. All repairs to the mScan™ must
be carried out by Myolex authorized technicians.
Transport
The mScan™ should always be transported in the case provided to ensure proper protection.
When being shipped, the mScan™ should be placed inside the case, ensuring a snug fit, and
the case should be placed in a shipping box with bubble wrap or shipping Styrofoam to
minimize movement within the box. A label should be placed on the outside of the box
indicating that its contents are fragile.
Troubleshooting
1. mScan™ Turns Off: If the mScan™ turns off, simply press the button on handheld
device to turn it back on. The mScan™ will turn off automatically if not used for
more than 3 minutes.