Users Manual

SENSE MCE IBC Operating Instruction
Rev. D, 2020-12-15 Release Page 25 of 46
© MYNXG Product GmbH, 2020. All rights reserved
3 Installation
IMPORTAN T NOTICE
The Sense MCE IBC shall not be installed on a place where it might be
penetrated by mechanical stress above 7 J.
CAUTION
Ensure that the location where the device is placed meets the specified
environmental conditions concerning temperature and humidity. not
cover the device and keep at least 20cm space around for air circulation.
CAUTION
The cable of the pre-assembled piezo is already connected to the Sense
MCE IBC. Ensure that the cable is handled carefully in order to prevent
any kind of damage.
IMPORTAN T N OTICE
Do not place the device close to high power / high frequency cable
tacks that might have an influence on the device minimum distance
50cm.
Do not place the device below ground level or inside rooms with
concrete walls (bunker like environments).
3.1 Delivery and Unpacking
The Sense MCE IBC is delivered in a carton enclosure which also includes the (optional)
(mounting plate. The device is already pre-assembled and personalized, i.e. ready for use after
waking up.
IMPORTAN T NOTICE
Please check packaging and delivery immediately after receipt regarding
damages and completeness. Do not throw away the packaging it can
be used for storage or sending the device back to MYNXG in case of
service needed.
CAUTION
Devices on which moisture has developed because of temperature
changes must not be used unless they have been completely dried.
CAUTION
The SENSE MCE IBC shall not be used when damages or technical
defects are visible or recognized.