User`s guide
Rockhopper 38 – User’s Guide
Page 135 AP-75038, Rev. 1.0. – 27/05/04
6.12.2 Error Messages
Message Action
Front L Sens Err
Error: Front Left Temperature Sensor
Front R Sens Err
Error: Front Right Temperature Sensor
Fixer L Sens Err
Error: Fixer Left Temperature Sensor
Fixer R Sens Err
Error: Fixer Right Temperature Sensor
Front Both Error
Error: Both Front Temperature Sensors
Fixer Both Error
Error: Both Fixer Temperature Sensors
Fro L, Fix L Err
Error: Front Left, Fixer Left Temperature Sensors
Fro R, Fix L Err
Error: Front Right, Fixer Left Temperature Sensors
Fro L, Fix R Err
Error: Front Left and Fixer Right Temperature Sensors
Fro R, Fix R Err
Error: Front Right and Fixer Right Temperature Sensors
Fr L&R, Fx L Err
Error: Both Front, Fixer Left Temperature Sensors
Fr L&R, Fx R Err
Error: Both Front, Fixer Right Temperature Sensors
Fr L, Fx L&R Err
Error: Front Left, Both Fixer Temperature Sensors
Fr R, Fx L&R Err
Error: Front Right, Both Fixer Temperature Sensors
All Sensor Error
Error: All 4 Temperature Sensors
AC Not Present
There is no AC current (heater strip could not heat )
Safety Cool Down
Wait..., Restart
In last runtime, there was some error, and
in this runtime, one of heater strips > 35 ºC.
You must let the machine cool down, then restart it.
Cool Down Done
Please Restart
If both strip temperature <35 ºC
(Above LCD text display will changed to this text)
Strips Testing
Could Not Be Run
Please Check
Sensors Connect
One or more sensors are not connected correctly. Please check the sensor
connection
Temperature
too high for
strip testing,
test is skipped.
One of heater strips > 35 ºC, heater strip testing (check) is skipped. If you
really want to test heater strip, please cool down machine, and restart it.
Front strip err
Front strip has connection error
Fixer strip err
Fixer strip has connection error
Front Strip> 65 ºC
Front heater has too high temperature
Fixer Strip > 85 ºC
Fixer heater has too high temperature
Group Front Err
Two sensors on Front strip have 20 ºC difference
Group Fixer Err
Two sensors on Fixer strip have 20 ºC difference
Front Fail Heat
Front strip can not reach target temperature within 10 minutes
Fixer Fail Heat
Fixer strip can not reach target temperature within 10 minutes
Both SSR Error
Both Solid State Relais have failed
Fixer SSR Error
Fixer Solid State Relais has failed
Front SSR Error
Front Solid State Relais has failed