Macros Tab
Conditional automatic command functions can be done with
macros. Click on the Macros tab to see available macros.
CALIBRATION
To calibrate a metering channel, double-click on the meter for that
channel in the Monitor window to open the Calibration window.
The Calibration window (shown below) allows modification of both
Upper and Lower Limits, the minimum and maximum views
displayed, and the calibration of the currently displayed value.
To change the calibration, click on the value field to edit the
displayed calibrated value. Once you have made the changes,
click on the Calibrate button instead of hitting enter. This will
change the displayed calibrated value to the new value. The view
fields will use the new calibration constant for all future input
values on this channel. When you are finished adjusting the limit
parameters and calibration, click OK or Apply to save changes
and return to the Monitor window.
The displayed calibrated value will not change on other PCs that
are connected at the same time. To see the new values on the
other connected PCs, the user must download the changes from
the unit or reconnect to the unit.
Meter Calibration Window
Always calibrate your
input when the sample is
at it’s peak level. If you
calibrate using a lower sample,
the calibrated value may change
dramatically once the peak value
is reached.
!