Specifications

MEA-System Manual
14
3 Extracellular Recording from MEAs
3.1 Introduction
Over the last 30 years, non-invasive extracellular recording from multiple electrodes has
developed into a widely used standard method. Systems and methods have been greatly
improved, leading to more features, higher throughput, and lower costs. Almost all excitable or
electrogenic cells and tissues can be used for extracellular recording in vitro, for example, central
or peripheral neurons, heart cells, retina, or muscle cells.
3.2 Background
The semi permeable lipid bilayer cell membrane separates different ion concentrations (charges)
on the inner and outer side of the membrane. Therefore, the cell membrane has the electrical
properties of a plate capacitor. The electrochemical gradient results in a membrane potential that
can be measured directly with an intracellular electrode. When ion channels are opened due to
chemical or electrical stimulation, the corresponding ions are moving along their electrochemical
gradient. In other words, the resistance of the membrane is lowered, resulting in an inward or
outward flow of ions, measured as a transmembrane current.
The extracellular space is conductive as well, and though the resistance is very low, it is not zero.
According to Ohm's law (U=R*I), the extracellular current results in a small voltage that can be
measured with extracellular electrodes. Extracellular signals are smaller than transmembrane
potentials, depending on the distance of the signal source to the electrode. Extracellular signal
amplitudes decrease with increasing distance of the signal source to the electrode. Therefore, a
close interface between electrode and cell membrane is very important for a high signal-to-noise
ratio.
The transmembrane current and the extracellular potential follow the same time course and are
roughly equal to the first derivative of the transmembrane potential.
A microelectrode array (MEA) is an arrangement of several (typically 60) electrodes allowing
targeting several sites for stimulation or recording at once.
The following components are important for an extracellular recording system:
 Signal source (cells / tissue)
 Cell / sensor interface
 Biosensor (MEA)
 Filter amplifier (MEA1060)
 Recording hardware (MC_Card) and software (MC_Rack)