Owner's manual

Temperature Controller TC01/02 Manual
20
4.3 MCS Default PI Coefficients
Note: The following PI parameters have been optimized at an ambient temperature of 25 °C,
the PI coefficients for use with the PH01 at a flow rate of 3 ml/min. You may have to adjust these
PI coefficients for your experimental setup, especially if the ambient temperature or the flow rate
differ at large from those used by MCS. Using suboptimal PI coefficients can lead to an oscillation
of the actual temperature, which is harmless, but may result in an unwanted behavior of the
temperature controller.
TCW1 PH01 MEA 1060-UP MEA 1060-INV
P 60.00 2.00 4.00 6.00
I 75.00 2.00 1.50 0.90
4.4 Technical Specifications
Operating temperature 10 °C to 40 °C
Storage temperature 0 °C to 50 °C
Dimensions (W x D x H) 170 mm x 224 mm x 66 mm
Weight 1.5 kg
Supply voltage and current 24 V and 4 A
Desktop AC Power Adapter 85 VAC to 264 VAC @ 47 HZ to 63 Hz
Sensor type PT 100
Measuring method four wire measuring bridge
Measuring temperature range 0 °C to 105 °C
Number of output channels 1 (TC01), 2 (TC02)
Output voltage max. 24 V
Output current max. 2.5 A per channel
Output power max. 30 W per channel
Resistance of heating element
5 - 100
Control range Ambient temperature (min. 5 °C) to 105 °C
Control interface USB 2.0
Note: Outputs are isolated against ground.