Instruction Manual
Appendix
85
Timing on external trigger
Minimum time length of trigger input (TTL) 40 ± 10 μs
Time lag between trigger input and Sync Out 30 ± 10 μs
Maximum trigger input frequency 10 kHz
Interface and connectors
Sync Out for synchronization with following devices 1 Lemo connector
Trigger In for synchronization with preceding devices 1 Lemo connector
Analog Out for analog channels 8 Lemo connectors
Interface (connection to computer) USB 2.0 High Speed
Filter Amplifier
Number of amplifier 8
Input voltage range ± 500 mV
Input impedance
>10
9
parallel to 10 pF
Input noise < 1 μV RMS (full bandwidth, inputs short circuited)
Noise density e
n
= 9 nV / rout Hz
Output voltage range ± 5 V
Output current 10 mA
Bandwidth 1 Hz to 2200 Hz
Gain options 100, 200, 500 or 1000
MC_Stimulus program
MC_Stimulus II Version 3.4.0 and higher
Operating system Microsoft Windows ® 7, XP or Vista with NTFS
English and German versions supported
Data import and export ASCII file format