13.1 Technical Specifications STG3008-FA
European Standard: EN 55011, EN 61000-6-1
General characteristics
Operating temperature 10 °C to 50 °C
Storage temperature 0 °C to 70 °C
Dimensions (W x D x H) 220 mm x 170 mm x 60 mm
Weight 1050 g
Fuse internal, 3.15 A slow blow
Supply voltage (external power supply) 100 to 240 VAC @ 47 to 63 Hz
Stimulus Generator
Number of stimulus generators 8
Output voltage range -8 V to + 8 V @ max. ± 20 mA
Output voltage resolution 1 mV
Rise time (10 % - 90 %) voltage 1.7 μs @ delta U = 8 V
Time lag between Sync Out and voltage
output
15 μs ± 2 μs for continuous pulses
40 μs ± 5 μs for a first or a single pulse
Output current -1.6 mA to +1.6 mA @ max.± 120 V compliance
voltage (between +I and I)
Output current resolution 200 nA
Output current slope
600 μA/μs @ R
L
= 10 k
Time lag between Sync Out and current
output, 0 - 100 μA
15 μs ± 2 μs @ R
L
= 10 k
Rise time (10 % - 90 %) current, 0 - 100 μA
1.4 μs @ R
L
= 10 k
Resolution 14 bit
Time resolution 20 μs
Output signals Freely programmable (rectangular, ramp, sinus,
arbitrary
Maximum frequency (rectangular waveform) 25 kHz
Memory capacity for stores waveforms 64 MB