Manual

Technical Specifications
Page 2 of 2 © 2011 Multi Channel Systems MCS GmbH
Timing on external trigger
Minimum time length of trigger input (TTL) 40 ± 10 μs
Time lag between trigger input and Sync Out 30 ± 10 μs
Maximum trigger input frequency 10 kHz
Interface and connectors
Sync Out for synchronization with following devices 1 Lemo connector
Trigger In for synchronization with preceding devices 1 Lemo connector
Analog out for analog channels 8 Lemo connectors
Interface (connection to computer) USB 2.0 High Speed
Download rate max. 480 Mb per second
Filter Amplifier
Number of amplifiers 8
Input voltage range ± 500 mV
Input impedance
> 10
9
parallel to 10 pF
Input noise < 1 μV RMS (full bandwidth, inputs short circuited)
Noise density
e
n
= 9 nV / Hz
Output voltage range ± 5 V
Output current 10 mA
Bandwidth 1 Hz to 2200 Hz
Gain options 100, 200, 500, or 1000
Software
MC_Stimulus II Version 3.4.0 and higher
Operating system Microsoft Windows ® 7, Vista or XP with NTFS;
English and German versions supported
Data import / export ASCII file format