System Manual User Manual
Appendix
57
Input Connectors of the MEA2100-System from Contact Unit to Headstage
(100-Pin Fischer Connectors 50 mil)
Connector on the left side, when looking to the front of the MEA2100 headstage
without contact unit and with the articulation of the headstage in the back.
Pin 1 and 2 GND (Ground)
Pin 3
to 30
not connected
Pin 31
to 34 internal coding
Pin 35
to 38
GND (Ground)
Pin 39
to 98
Channel 1 to 60
Pin 99 and 100 GND (Ground)
Connector on the right side, when looking to the front of the MEA2100 headstage
without contact unit and with the articulation of the headstage in the back.
Pin 1 and 2 GND (Ground)
Pin 3 and 62 Channel 61 to 120
Pin 63
to 66 GND (Ground)
Pin 67
to 70
internal coding
Pin 71
to 98 not connected
Pin 99 and 100 GND (Ground)