60StimMEA
60StimMEA200/30-ITO
60StimMEA200/30-Ti
Technical Specifications 60 Electrode MEA with 16 additional Stimulation Electrodes
Temperature compatibility 0 - 125 °C
Dimension (W x D x H) 49 mm x 49 mm x 1 mm
Base material Glass
Track material Indium tin oxide (ITO) or Titanium (Ti)
Contact pads Indium tin oxide (ITO) or Titanium nitride (TiN)
Electrode diameter 30 m
Interelectrode distance (centre to centre) 200 m
Electrode height Planar
Electrode type Titanium nitride (TiN) electrodes
Isolation type Silicon nitride 500 nm (PEVCD)
Electrode impedance 30 - 50 k
Electrode layout grid 8 x 8 and additional
4 pairs of large stimulation electrodes (STIM = 70 x 250 m)
4 pairs of small stimulation electrodes (S = 30 m)
Number of recording electrodes 60
Number of reference electrodes Without internal reference electrode
MC_Rack
Source layout in “Data Source Setup” 2 dim. (MEA)
Channel map Default
MEA perfusion chamber (w/o) without ring
(gr) Glass ring: ID +/-19 mm, OD 24 mm, height 6 / 12 mm
60 electrodes Microelectrode Array
with 16 additional stimulation electrodes.
Electrode layout: Standard 8 x 8 grid
For use with MEA-STIM-ADPT:
Adapter for MEAs with 16 additional
stimulation electrodes and for
MEA1060 amplifiers.
21 31
41
51 61
71
12
13
14
15
16
17
22
23
24
25
26
27
28
32
33
34
35
36
37
38
42
43
44
45
46
47
48
52
53
54
55
56
57
58
62
63
64
65
66
67
68
72
73
74
75
76
77
78
82
83
84
85
86
87
33 21 32 31 44
43
41 42 52 51
53
54
61
62
71
63
22
12
23
13
34
24
14
15
25
35
16
26
17
27
36
28
37 38
45 46 48 47 57 58
56 55
68
67
78
66
77
87
76
86
72
82
73
83
64
74
84
85
75
65
S
S
S
S
S
S
S
S
STIM
STIM
STIM
STIM
STIM
STIM
STIM
STIM
2450 μm
250 μm
7,755 mm
4,1 mm
7,755 mm
27,8 mm
Contact pad for stimulation adapter
MEA-STIM-ADPT
Diameter of contact pad 2 mm
Distance between contact pads 4.1 mm
(Centre to centre)
2000 μm