60ThinMEA
60ThinMEA30/10-ITO
60ThinMEA100/10-ITO
60ThinMEA200/30iR-ITO
Technical Specifications 60ThinMEA
Temperature compatibility 0 - 125 °C
Dimension (W x D x H) 49 mm x 49 mm x 1 mm
“Thickness” 180 m (Glass part)
Base material Glass on ceramic carrier
Contact pads and track material ITO (Indium tin oxide)
Electrode diameter 10 or 30 m
Interelectrode distance (centre to centre) 100 or 200 m
Electrode height Planar
Electrode type TiN (Titanium nitride)
Isolation type Silicon nitride 500 nm (PEVCD)
Electrode impedance 30 - 50 k for 30 m, 250 - 400 k for 10 m electrodes
Electrode layout grid 8 x 8
Number of recording electrodes 59 (with iR) or 60 (without iR)
Number of reference electrodes 1 internal reference electrode iR: 60ThinMEA200/30iR-ITO
or without internal reference electrode: 60ThinMEA100/10-ITO
60ThinMEA30/10-ITO
MC_Rack
Source layout in “Data Source Setup” 2 dim. (MEA)
Channel map Default
MEA perfusion chamber (w/o) Without ring
(gr) Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm
(pr) Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm
(pr-T) Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm
21 31
41
51
61
71
12
13
14
15
16
17
22
23
24
25
26
27
28
32
33
34
35
36
37
38
42
43
44
45
46
47
48
52
53
54
55
56
57
58
62
63
64
65
66
67
68
72
73
74
75
76
77
78
82
83
84
85
86
87
33 21 32 31 44
43
41 42 52 51
53
54
61
62
71
63
22
12
23
13
34
24
14
15
25
35
16
26
17
27
36
28
37 38
45 46 48 47 57 58
56 55
68
67
78
66
77
87
76
86
72
82
73
83
64
74
84
85
75
65
60 electrode Thin Microelectrode Array with 8 x 8 layout.
The electrodes are embedded in a very thin glass substrate
on a robust ceramic carrier. Contact pads and tracks are made
from transparent indium tin oxide for high resolution imaging.