6.1 Technical Specifications
The MEA-IT-System is a device for impedance testing of microelectrode arrays.
Warning: The device may only be used together with microelectrode arrays from
Multi Channel Systems MCS GmbH, and only for the specified purpose. Damage of
the device and even fatal injuries can result from improper use. Do not open the data
acquisition box and do not change hardware configuration as it could lead to improper
behavior of the system.
General characteristics
Operating temperature 10 °C to 50 °C
Storage temperature 0 °C to 50 °C
Relative humidity 10 % to 85 %, non-condensing
Dimensions (W x D x H) 250 mm x 140 mm x 25 mm
Weight 1100 g
Number of channels 60
Measurement range
5 k to 2 M
Test signal 100 mV, 1 kHz Sinus
Accuracy 5 %
Power supply USB powered
PC interface USB 2.0
Software
Operating system Microsoft Windows ® 7 (32 or 64 bit), XP or Vista
with NTFS, English and German version supported
MEA-IT program Version 1.0.0 and higher