Technical Specifications FlexMEA72
Temperature compatibility 10 - 40 °C
Dimension (W x D) of the FlexMEA72 42 mm x 12 mm
Thickness of the electrode field 12 m
Weight < 1 g
Base material Polyimide 2611 foil
Contact pads and track material Gold
Electrode diameter 100 m
Interelectrode distance (centre to centre) 625 to 750 m
Perforation Diameters of the holes 100 m
Electrode height Planar
Electrode type TiN electrodes (Titanium nitride)
Isolation type Polyimide 2611 foil
Electrode impedance Approximately 50 k
Electrode layout grid 9 x 8
Number of recording electrodes 64
Number of reference electrodes 4 internal reference electrodes
Number of ground electrodes 4
MC_Rack Version 4.1.1 and higher
Data Source Setup Source Layout: Configuration, Total Number of Channels: 64,
Amplifier: FA64I/S, MEA: FlexMEA72.
Channel map Display: Tab Layout > Default Map (FlexMEA72.cmp).
Cleaning Rinse with distilled water, optional with ethanol 70%.
Do not autoclave or sterilize FlexMEAs by heat. These MEA types are not heat-stable
and will be irreversibly damaged! Please do not use an ultrasonic bath for cleaning.
Flexible Microelectrode Array with 72 electrodes for use via ADPT-FM-72 adapter with
two 32-Channel Miniature Preamplifier MPA32I for in vivo and in vitro applications.