Owner's manual
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25-
0
Fax
+49-7121-9 09 25-11
info@multichannelsystems.com
www.multichannelsystems.com
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
120MEA-SG
120MEA Signal Generator for use with MEA2100-HS120-System
Control button
LED
DIP switch
Battery
Gold contact pads
120MEA Signal Generator is a convenient tool for MEA2100-System users. Use the 120MEA-SG
instead of setting up an experiment with biological sample for training, controlling, and
troubleshooting purposes. This reduces the number of animal experiments and saves laboratory
equipment.
Switch on : Press control button. Switch off : Press control button longer than two seconds.
Important : To change DIP switch position, please switch off the device!
!
Table : DIP switch positions, number of control button presses, and corresponding signals
Switch 1 Switch 2 Control button Signal
presses n times
OFF OFF MEA-SG ON Sinus 0.005 Hz
1 Sinus 0.01 Hz
2 Sinus 0.03 Hz
3 Sinus 1.25 Hz
4 Sinus 12.5 Hz
ON OFF MEA-SG ON EPSP
1 Population Spike
2 Spikes
OFF ON MEA-SG ON ECG Atrium
1 ECG Ventricle
2 Ventricle FP
ON ON MEA-SG ON ERG with Spikes