User Guide

17
Preparing The Unit (con.)
Adjust the angle of the sternum
support to a position similar to Figure
17. Use the allen wrench to loosen/
tighten the four screws. The angle of
the sternum pad can be adjusted further
once the user is in the unit. Do not
place the user in the unit at this time!
Adjust the distance between the sternum pad and the seat
so it is equal to measurement #2 as shown in Figure 18 by
moving the seat frame up/down on the center column.
Confirm that the seat height can be reached. Actuate the
gas spring so the upper frame rises until it comes to a stop.
The gas spring has a
7.9in (20cm) stroke.
Measurement #1
should be in between
the value of #5 (Figure
18) and #5
- 7.9in.
#5 #1 (#5
7.9in)
If this relationship is
not true, adjust the
location of the center
mast on the upper
frame using the screws
shown in Figure 18
until the relationship is
valid.
Figure 17
Screws
#5
#2
Figure 18
screws