User guide

Load becomes greater than 1.00 kN (Termination Condition)
Specimen failure is detected
One of the other branches of the Parallel Branch activity (such as a limit detection branch) completes
The Brake Distance control provides control-mode braking, which slows the command to help avoid
overshoot. The brake value is subtracted from the Termination Condition value to determine where
control-mode braking starts. In this case, the Termination Condition is 1.00 kN and the Brake Distance
is set to 100 N, at 900 N the command slows to 10% of the Test Rate (0.3 mm/s), and at 990 N, command
slows to 1% of the Test Rate (0.03 mm/s).
Modifying the Finish Section
Requirements
The test needs to automatically generate a test report and send it to a manager via e-mail when 24
test runs have been performed.
Modifications to the Finish section
Change the Maximum Test Runs variable to 24.
Enable the Run Report activity and select the Send report in e-mail option.
Changing the Maximum Test Runs Variable
1. Click the Variables subtab (Define tab > Variables subtab).
2. Select the Maximum Test Runs variable from the list of available variables and, in the Properties Panel,
change the default value to 24.
92 | MTS TestSuite