Owner manual

Use the following parameters:
Meter Type Timed
Channel Ch1 (or other depending on your configuration)
Signal Force
Engineering Units kN
Zero Displacement and Force Signals
With the actuator at mid-travel and no force on the load cell, zero the displacement and force signals.
1. Auto offset the Displacement signal.
a) Click the Station Setup button.
b) Select Channels > Ch1 > Displacement > Channel Input Signals button > Offset/Zero tab.
c) Click the Auto Offset button.
2. Auto offset the Force signal.
80 | MTS TestLine Test System Operation
Running the Example HCF Test