Manual

Routine Maintenance
Series FEC Environmental Chamber Product Manual Routine Maintenance
43
Routine Maintenance
Routine Maintenance Overview Checklist
Recommended Service
CALENDAR TIME USING 8 HOURS RUNNING TIME
R
ATE PER DAY
DAILY WEEKLY ANNUALLY
RUNNING TIME-HOURS 8 40 2000
Clean the chamber inside/chamber surface area
X
*
* Denotes services performed by equipment operators. Most of these procedures involve visual checks
that should not interfere with testing system operation. These checks are also completed by trained
field service engineers on each Routine Maintenance visit.
Check the door limit and E-stop X
Check the cable connection X
Check all the buttons function X
System Inspection
Check the cable connection MTS
System Checks
Check E-stop MTS
Check door switch MTS
Check rail bracket level MTS
Check over-temperature alarm instrument
function
MTS
Lubrication
Stirring fan bearing MTS