User Manual
Table Of Contents
- Technical Support
- Preface
- Introduction
- Overview Reference
- Series 647 Hydraulic Wedge Grips Component Identification
- Series 647 Hydraulic Wedge Grips Functional Description
- About Gripping Specimens
- About Wedges
- About Spiral Washers
- About Couplings
- About All Temperature Grips
- About Environmental Chambers
- Hydraulic Fluid Recommendations
- Cooling Water Specifications
- Series 647 Wedge Grip Temperature Ranges
- Series 647 Hydraulic Wedge Grip Force and Torque Capacities
- Safety
- General Safety Practices: Grips and Fixtures
- Read all manuals
- Avoid Pinch and Crush Points
- Locate and read hazard placards/labels
- Know facility safe procedures
- Know controls
- Know Specimen Properties
- Have first aid available
- Be aware of component movement with hydraulics off
- Keep bystanders safely away
- Wear proper clothing
- Remove flammable fluids
- Check bolt ratings and torques
- Lift Equipment Safely
- Practice good housekeeping
- Do not exceed the Maximum Supply Pressure
- Do not disable safety devices
- Provide adequate lighting
- Provide means to access out-of-reach components
- Wear appropriate personal protection
- Handle chemicals safely
- Know system interlocks
- Know system limits
- Do not disturb sensors
- Ensure secure cables
- Stay alert
- Contain small leaks
- Stay clear of moving equipment/avoid crush points
- Know the causes of unexpected actuator motions
- General Precautions for Environmental Components
- Hazard Placard Placement
- General Safety Practices: Grips and Fixtures
- Installation
- Operation
- Maintenance

DescriptionItemDescriptionItem
Lower Hydraulic Chamber6Piston Extension3
b) Lubricate the threads and the interior wall of the cylinder.
c) Position the cylinder assembly of the lower grip into the environmental chamber. Move the cylinder
assembly through the hole in the environmental chamber. Align it with the threads of the hydraulic
chamber.
d) Rotate the cylinder until you feel the thread ends engage. Hand-tighten the cylinder to the hydraulic
chamber until it is snug.
e) Lubricate the entire lower and bottom ends of the piston extension. Also lubricate the threads of the
socket head bolt. Insert the socket head bolt into the piston extension.
f) Slide the piston extension into the cylinder (narrow end first).
g) Thread the socket head bolt into the hydraulic chamber. Do not tighten it at this time.
h) Repeat Step A through Step G for the upper cylinder assembly and piston extension.
Note:
The piston might slide out of the cylinder when the cylinder end is pointed down.
7. Install the lower wedge chamber.
a) Install the eye bolts on the side of the wedge chamber.
DescriptionItemDescriptionItem
Cylinder Assembly4Upper Hydraulic Chamber1
Lower Hydraulic Chamber5Cylinder Assembly2
62 Series 647 Hydraulic Wedge Grips Reference Manual
Installation