User Manual
Table Of Contents
- Technical Support
- Preface
- Introduction
- Overview Reference
- Series 647 Hydraulic Wedge Grips Component Identification
- Series 647 Hydraulic Wedge Grips Functional Description
- About Gripping Specimens
- About Wedges
- About Spiral Washers
- About Couplings
- About All Temperature Grips
- About Environmental Chambers
- Hydraulic Fluid Recommendations
- Cooling Water Specifications
- Series 647 Wedge Grip Temperature Ranges
- Series 647 Hydraulic Wedge Grip Force and Torque Capacities
- Safety
- General Safety Practices: Grips and Fixtures
- Read all manuals
- Avoid Pinch and Crush Points
- Locate and read hazard placards/labels
- Know facility safe procedures
- Know controls
- Know Specimen Properties
- Have first aid available
- Be aware of component movement with hydraulics off
- Keep bystanders safely away
- Wear proper clothing
- Remove flammable fluids
- Check bolt ratings and torques
- Lift Equipment Safely
- Practice good housekeeping
- Do not exceed the Maximum Supply Pressure
- Do not disable safety devices
- Provide adequate lighting
- Provide means to access out-of-reach components
- Wear appropriate personal protection
- Handle chemicals safely
- Know system interlocks
- Know system limits
- Do not disturb sensors
- Ensure secure cables
- Stay alert
- Contain small leaks
- Stay clear of moving equipment/avoid crush points
- Know the causes of unexpected actuator motions
- General Precautions for Environmental Components
- Hazard Placard Placement
- General Safety Practices: Grips and Fixtures
- Installation
- Operation
- Maintenance
Turn the hydraulic pressure off. Ensure that system hydraulic pressure has been reduced to zero before
proceeding. To do this, turn off the hydraulic power unit and exercise the actuator until it stops moving.
Using the provided hydraulic hoses, make connections between the pressure and release ports on the
grip and the pressure and release ports on the load unit or on the external hydraulic grip supply. See your
load unit product manual or hydraulic grip control manual for their port locations.
6. If needed, connect the water supply. Water can heat or cool the grips.
a) Mount the water valve clamp to a load unit column.
b) Connect water lines (supply and drain) between the water lines attached to the columns and the upper
and lower grips.
c) Connect the water supply (and drain) to the water valve/drain mounted to the load unit columns.
7. Tighten the grips.
See “Preload the Spiral Washers” to preload the connector stud. Return to this procedure when done.
8. Install the environmental chamber.
Position the environmental chamber (with its cart) in the load unit so that the actuator extends through the
chamber’s U-plugs. Secure the cart and install the U-plugs. Refer to the environmental chamber
documentation for more information.
9. Install the wedges.
Go to “Change Wedges 647.02 - 647.100” for the procedure to install wedges into the grips.
Install Environmental Chambers Without U-Plugs
Environmental chambers that do not have U-plugs must be installed before the grips are installed. The grips
must be partially disassembled to install them. The following figure shows the components of the grip assembly.
Series 647 Hydraulic Wedge Grips Reference Manual 55
Installation