Owner's manual
Flat Thick Specimens
Series 609 Alignment Fixture Product Information Specimen Preparation
37
Flat Thick Specimens
A typical gaged specimen has three levels of strain gages. The gages allow you to
determine bending strains at each level. Analysis of the strain readings produces
a profile of the grip alignment.
Gage placement Use twelve gages on thick flat specimens. Attach and number them as shown
here.
Gage Placement—Thick Flat Specimens
• Place the gages in sets of four, at the top, middle, and bottom of the gage
section. Center each gage between the edges of the specimen.
• Place the top set of gages so that their upper edges just touch the upper edge
of the gage section.
• Place the middle set of gages in the middle of the gage section.
• Place the bottom set of gages so that their lower edges just touch the lower
edge of the gage section.
1
2
34
5678
9101112
0° 90° 180° 270°