Interlock logic
The interlock logic is as follows:
Interlock Logic
DescriptionItem
Door Interlock1
Solenoid Contacts2
Test Controller Program Interlock Input. Open = Interlock.3
To Low Flow Power Supply4
To Controller Program Interlock Chain5
Test Area Enclosure with Light Curtain
This section provides information on the test area enclosure with light curtain provided with some configurations
of the Series 370 Tabletop Load Unit.
The light curtain replaces the door typically included with a test area enclosure. This configuration of the test
area enclosure is typically used in QC elastomer applications. It includes operator control panels that require
both hands outside the test area to start the test. If the actuator is moving and the light curtain beam is
interrupted, the actuator motion stops until the beam is no longer interrupted.
70 MTS Landmark⢠Tabletop Load Units Product Information