Landmark Test System Site Prep Guide
16
Load Unit Specifications
System Component Specifications
Environmental Requirements—Series 370 Load Frames
All Series 370 load frames are intended for indoor use only. This indoor
environment must conform to the following environmental specifications.
Note All Series 370 load frames must only be operated under the installation
and ambient conditions (such as, temperature, moisture, and EMC)
specified.
Electrical Requirements—Series 370 Load Frames
Parameter Specification
Temperature 5ºC–40ºC (41ºF–104ºF)
Humidity 10–85%, non-condensing
Altitude 2000 m (6,561 ft) maximum
Parameter Specification
Input Voltage 115 or 230 VAC (Single phase)
Input Frequency 50 or 60 Hz
Input Current 7 A at 115 VAC, 0.4 A at 230 VAC
Environment: Location: Indoor use only
Operating temperature 5–40°C (41–104°F)
Relative humidity 5-85% non-condensing
Maximum altitude 3000 m (9,842 ft)