User guide
MTS Criterion™ Series 60
40
Appendix
Foundation Requirements
• Cast the foundation according to the above drawings and dimensions.
• The compression stress of the concrete used for the foundation should be no
less than 35 Mpa (5000 psi).
• Refer to the previous table for anchor bolts. The tensile stress of the bolts
should be no less than 450 MPa (64286 psi). Tighten the bolts with cushion
rings, spring cushions, and nuts on the frame base. After the tightening, the
bolts should extend 5-10 mm (0.2-0.4 in) above the nuts.
• While recasting the reserved holes, keep the anchor bolts at the centers of
the holes.
• Recast the reserved holes to the same plane of the floor. Cure the foundation
until it is qualified for use.
• The degree of the foundation level should be no more than 0.2/1000.
Note The above foundation drawings are for reference only; you can make
alternations according to your actual needs.
Foundation Board Measurement
64.305EL 64.605/64.605EL 64.106/64.106EL
Ⅰ
(L×W×T)
840×828×50 mm
(33×32.6×2 in)
996×829×50 mm
(39.2×32.6×2 in)
1126×825×50 mm
(44.3×32.4×2 in)
Ⅱ , Ⅳ
(L×W×T)
2305×840×50
mm
(90.7×33×2 in)
2330×840×50 mm
(91.7×33×2 in)
2440×840×50 mm
(96.2×33×2 in)
Ⅲ
(L×W×T)
840×852×50 mm
(33×33.5×2 in)
996×855×50 mm
(39.2×33.6×2 in)
1126×858×50 mm
(44.3×33.7×2 in)
a
(Gap)
115 mm
(4.5 in)
163 mm
(6.4 in)
168 mm
(6.6 in)
b
(Gap)
812 mm
(32.0 in)
815 mm
(32.0 in)
818 mm
(32.0 in)
Ⅴ
[Rect. tube of 5
mm (0.2 in)
thickness]
140×80×2580
mm
(5.5×3.3×101.7
in)
140×80×2736 mm
(5.5×3.3×107.8 in)
140×80×2866 mm
(5.5×3.3×113.0 in)