Instruction Manual
Controller Connections
MTS Criterion™ Series 60 Product Manual Installation
69
J11 Guard This connector is intended to connect to a test area enclosure switch that opens
when the door is opened. C60 series do not use this function, a jumper plug must
be installed. A jumper plug must be installed. If you are building a cable, the
maximum length is 30.48 m (100 ft) with 22 AWG. The switch should be wired
normally closed, such that when the switch opens, an interlock is generated. Pin
assignment is as follows:
GUARD = test area guard switch with two contacts.
GUARD_CONFIG = configuration jumpers in the cable.For speed limit when
test are guard door open jumper Guard_config1+ to Guard_config1- and
Guard_config2+ to Guard_config2-. For interlock when guard door open leave
Guard_config unconnected.
J12 Digital I/O The connector is the MTS Insight digital I/O control connector. It is connected
with J5
Digital I/O connector through a jumper.
J13 Digital I/O User Digital I/O signals include three optically isolated inputs, three optically isolated
outputs, and 12 V power. Functions of each digital input or output are software
selectable. A typical example might be connecting an external switch; see
“Additional Digital I/O Information” on page 93. Pin assignment is as follows:
Pin Assignment
PIN SIGNAL
1 GUARD 1A+
2 GUARD2A+
3 GUARD_CONFIG1+
4 No Contact
5 GUARD_CONFIG2-
6 GUARD1A -
7 GUARD2A-
8 GUARD_CONFIG1
9 GUARD_CONFIG2+
Pin Assignment
PIN SIGNAL
1 DIN 1+
2 DIN 2+
3 DIN 3+
4 OUT 1+
5 OUT 2+
6 OUT 3+