Instruction Manual
MTS Criterion™ Series 60 Product Manual
68
Controller Connections
Installation
J8 and J9 DC
Conditioner
Two DC Conditioner connectors are provided. The two application-specific
transducers might be biaxial strain gage base extensometers. With external
completion resistors, the DC conditioners could be used with quarter bridge
strain gages. Pin assignment is as follows:
J10 Aux Pwr This is an auxiliary power connector. It can provide power supply to
external devices. Pin assignment is as follows:
Pin Assignment
PIN SIGNAL
1 TEDS data
2 EX+
3 EX-
4 FB-
5 RCAL1 (FBR+)
6 RCAL2 (FBR-)
7 FB+
8 EXS-
9 EXS+
10 TEDS ground
Pin Assignment
PIN SIGNAL
1 +12V (Combined with
Pin 6
2 AGND
3 -12V
4 PGND
5 +5V
6 +12V ( Combined with
Pin 1)
7 AGND
8 DGND
9 +24V