Controller Connections
Installation
J4 Encoder This connector is intended for an encoder-based extensometer. Specifics for this
connector are:
Power: +5 V +/- 0.25 V at 100 mA max
Signals: Quadrature A and B with index I
Logic: Differential receivers (can connect single ended)
Maximum Rate: 100,000 lines/sec = 400,000 counts/sec
Pin assignment is as follows:
Pin Assignment
PIN SIGNAL
1 TEDS data
2 A+
3 A-
4 +5 V
5 I+
6 I-
7 Analog GND
8 B+
9 B-
10 TEDS ground