Instruction Manual
MTS Criterion™ Series 60 Product Manual
24
Specifications
Introduction
Specifications – Integrated Operations Platform
The following integrated operations platform specifications are common for
models of 64.305/64.305E, 64.605/64.605E, and 64.106/64.106E.
Integrated Operations Platform Specifications
PARAMETER SPECIFICATION
Calibration standard ISO 7500 Class 1 or ASTM E4
Force range 1%-100% of rated force capacity
Force indicating accuracy ± 1% of indicating
Position resolution 0.2 μm
(0.00000787 in)
Position accuracy +/- 1% of indicating
Strain accuracy ASTM E83 or ISO 8513
Security protection over-force, travel limits, over-pressure, over-
temperature, over-current and others
Over force protection 10%
Data acquisition rate 1000 Hz maximum
Control loop rate 1000 Hz
Environmental requirements
Operating temperature range 5-40°C (41-104°F)
Operating humidity 5-85% noncondensing
Storage temperature range 18-49°C (0 -120°F)
Maximum storage humidity 90% noncondensing
Maximum altitude 2000 m (6500 ft)
Integrated Operations Platform power
requirements
3~3W+PE
400 V 50 Hz / 480 V 60Hz
Integrated Operations Platform dimensions
Height 1900 mm
(74.80 in)
Width 1040 mm
(40.94 in)