Installing Cables
Installation
J14 E- Stop This connector is intended to connect to a remote Emergency-Stop switch. If not
used, a jumper plug must be installed. If you are building a cable, the maximum
length is 30.48 m (100 ft) with 18 AWG. The switch should be wired normally
closed, such that when the switch is pressed, an interlock is generated. Pin
assignments are as follows:
ESTOP3 = emergency stop switch with two sets of contacts (optional
remotely mounted).
ESTOP_OUT = optional ESTOP output for accessory, includes two contact
outputs and one feedback. For use with pneumatic/hydraulic grip controls,
furnaces, environmental chambers, and so on.
P
IN SIGNAL
1 ESTOPB_OUT-
2 No Contact
3 ESTOP_OUT_MONITOR-
4 ESTOP_OUT_MONITOR+
5 ESTOP3B+
6 ESTOPA_OUT-
7 ESTOP3B-
8 ESTOP3A+
9 ESTOPB_OUT+
10 No Contact
11 No Contact
12 No Contact
13 ESTOP3A-
14 ESTOPA_OUT+
15 No Contact