Owner's manual
Electrical connections must be made by qualified personnel and conform to local codes and regulations.
Local electrical codes supersede any information found here.
MTS Acumen 3MTS Acumen 1Electrical Requirements
200-240 VAC100-120 VAC (200-240)Voltage
50-60 Hz50-60 HzFrequency
10 Amp7 (4) AmpCurrent
12
SingleSinglePhase
Environmental Requirements
All MTS Acumen load frames are intended for indoor use only. This indoor environment must conform to the
following environmental specifications.
Environmental Requirements
SpecificationParameter
5° to 40° C (41° to 104° F)Operating Temperature
13
5 to 85% Non-condensingOperating Humidity
-18° to 49° C (0° to 120° F)Storage Temperature
90% Non-condensingMaximum Storage Humidity
2000 m (6562 ft)Maximum Altitude
2Pollution Degree
IP20IP Rating
IIOvervoltage Category
12
MTS Acumen 1 current rated at 100 VAC. MTS Acumen 3 current rated at 200 VAC.
13
Take care when using heat producing accessories, such as ovens. Emitted heat from such devices may
damage the MTS Acumen if not properly controlled.
MTS Acumen™ Electrodynamic Test System 29
System Specifications