3.8 MISCELLANEOUS PRECAUTIONS
The following are miscellaneous precautions associated with MTD System operation.
When necessary, the transfer line should be evacuated with a nitrogen-trapped diffusive-type
vacuum system.
The transfer line should not be bent to less than a 12-inch (30 cm) radius.
Heater power should not exceed 80 watts (first or second stage on cryostat).
The DUT mount or any part of the MTD-150 should never be heated above 100 °C.
Do not “break” vacuum in the chamber while the first or second stage is cold.
Do not “break” transfer line vacuum with helium gas.
Avoid contact with cold gas when depressurizing the storage dewar.
Periodically, the o-ring seals that are routinely accessed and their mating surfaces on the
MTD-150 should be cleaned with a paper wipe and re-greased with Apiezon
®