Instruction manual
Mettler Electronics Corp. — Rev.M_11/19/14
26
6. Press the stop button to stop treatment and release the traction force. The system will revert
to Normal Setting status.
NOTE: The patient can press the “Patient Safety Button” to stop treatment and release the
traction force if any discomfort is experienced during the treatment. The alarm buzzer will sound
and the display will show an error code “E-4”. To revert to a safe status “E-4” can be removed by
pressing the stop button.
6.3 Setting up the Intermittent Mode
Figure 6.2—Intermittent Mode Display
1. Press the output mode button twice to set the MTD 4000 to the intermittent mode. The LED
next to
should be lit.
2. Press the set button to set traction force. The “Hold” display will start flashing. Turn the
control knob to set the desired force value. Press the set button again to save the Hold force
value. Then the “Rest” display will start flashing. Turn the control knob to set the desired rest
force value. Press the set button again to save the rest force value. NOTE: When an
indicator is flashing, no key press within 10 seconds will cause the flashing to automatically
stop.
3. Press the time button to set the treatment time. The timer display will start flashing. Turn the
control knob to set the desired time in minutes. Press timer button again to save the time.
Flashing stops once the time is saved.
4. Press the hold button to set the hold time. The hold display will start flashing. Turn the
control knob to set the desired hold time in seconds. Press hold button again to save the
hold time. Flashing stops once the hold time is saved.
5. Press the rest button to set the rest time. The rest display will start flashing. Turn the control
knob to set the desired rest time in seconds. Press timer button again to save the rest time.
Flashing stops once the rest time is saved.
6. Press the speed button to set traction speed (FAST/NORMAL/SLO
W
).