Instruction manual
Mettler Electronics Corp. — Rev.M_11/19/14
22
6. Push platform to opposite side.
7. Position and tighten bolts.
8. Connect the line cord to the back of the MTD 4000. (See Figure 5.2)
9. Plug the line cord into a grounded wall outlet that is rated at AC 110V~120V, 50/60Hz (other voltages
available). Your mains power supply must match the voltage requirements listed on the serial number
label of your device.
10. Do not connect the MTD 4000 to a power supply rated differently than that described above. The
unit comes equipped with a grounded line cord. This plug provides grounding for the MTD 4000. Do not
defeat its purpose by using adapters or any other means of attaching to a wall outlet.
Figure 5.2— Rear View of the MTD 4000 Showing Connectors and Switches
11. Plug the patient safety switch into the back of the unit.
12. Attach the desired harness for either lumbar or cervical traction.